Utah Administrative Code (Current through November 1, 2019) |
R313. Environmental Quality, Waste Management and Radiation Control, Radiation |
R313-28. Use of X-Rays in the Healing Arts |
R313-28-80. Intraoral Dental Radiographic Systems
-
In addition to the provisions of R313-28-31, R313-28-32 and R313-28-35, the requirements of this section apply to x-ray equipment and associated facilities used for dental radiography. Criteria for extraoral dental radiographic systems are covered in R313-28-51, R313-28-52 and R313-28-53. Intraoral dental radiographic systems used must meet the requirements of R313-28-80.
(1) Source-to-Skin distance (SSD). X-ray systems designed for use with an intraoral image receptor shall be provided with means to limit source-to-skin distance to not less than:
(a) 18 centimeters if operable above 50 kilovolts peak, or
(b) 10 centimeters if not operable above 50 kilovolts peak.
(2) Field limitation. Radiographic systems designed for use with an intraoral image receptor shall be provided with means to limit the x-ray field so that:
(a) if the minimum source-to-skin distance (SSD) is 18 centimeters or more, the x-ray field, at the minimum SSD, shall be containable in a circle having a diameter of no more than seven centimeters; and
(b) if the minimum SSD is less than 18 centimeters, the x-ray field, at the minimum SSD, shall be containable in a circle having a diameter of no more than six centimeters.
(3) Exposure Initiation.
(a) Means shall be provided to initiate the radiation exposure by a deliberate action on the part of the operator, for example, the depression of a switch. Radiation exposure shall not be initiated without a deliberate action; and
(b) It shall not be possible to make an exposure when the timer is set to a "zero" or "off" position if either position is provided.
(4) Exposure Termination.
(a) Means shall be provided to terminate the exposure at a preset time interval, preset product of current and time, a preset number of pulses, or a preset radiation exposure to the image receptor.
(b) An x-ray exposure control shall be incorporated into x-ray systems so that an exposure of more than 0.5 seconds can be terminated immediately by the operator.
(c) Termination of an exposure shall cause automatic resetting of the timer to its initial setting or to "zero."
(5) Exposure Indication. Means shall be provided for visual indication, observable from the operator's protected position, whenever x-rays are produced. In addition, a signal audible to the operator shall indicate that the exposure has terminated.
(6) Timer Linearity. For systems having independent selection of exposure time settings, the average ratio of exposure to the indicated milliampere-seconds product obtained at two consecutive timer settings or at two settings not differing by more than a factor of two shall not differ by more than 0.10 times their sum.
(7) Exposure Control Location and Operator Protection.
(a) Stationary x-ray systems shall be required to have the x-ray exposure control mounted in a protected area or a means to allow the operator to be at least 2.7 meters (9.0 feet) from the tube housing assembly while making exposures; and
(b) Mobile and portable x-ray systems which are:
(i) used for greater than one week in the same location, for example, a room or suite, shall meet the requirements of R313-28-80(7)(a); or
(ii) used for less than one week in the same location shall be provided with either a protective barrier at least two meters high for operator protection, or means to allow the operator to be at least 2.7 meters (nine feet) from the tube housing assembly while making exposures.
(8) Exposure Reproducibility. When all technique factors are held constant, the coefficient of variation of exposure shall not exceed 0.05 for certified x-ray systems or 0.10 for non-certified x-ray systems. This requirement applies to clinically used techniques.
(9) mA/mAs Linearity. The following requirements apply when the equipment is operated on a power supply as specified by the manufacturer for fixed x-ray tube potentials within the range of 40 to 100 percent of the maximum rated potentials.
(a) For equipment having independent selection of x-ray tube current, the average ratios of exposure to the indicated milliampere-seconds product obtained at two consecutive tube current settings or, when the tube current selection is continuous, two settings differing by no more than a factor of two shall not differ by more than 0.10 times their sum.
(b) For equipment having a combined x-ray tube current-exposure time product selector but not a separate tube current selector, the average ratios of exposure to the indicated milliampere-seconds product obtained at two consecutive mAs selector settings, or when the mAs selector provides continuous selection, at two settings differing by no more than a factor of two shall not differ by more than 0.10 times their sum.
(10) Accuracy. Deviation of technique factors from indicated values shall not exceed the limits specified for that system by its manufacturer. In the absence of manufacturer's specifications the deviation shall not exceed ten percent of the indicated value.
(11) Administrative Controls.
(a) Patient and film holding devices shall be used when the technique permits and holding is required.
(b) The x-ray tube housing and the position indicating device shall not be hand-held during an exposure.
(c) The x-ray system shall be operated so that the useful beam at the patient's skin does not exceed the requirements of R313-28-80(2).
(d) Dental fluoroscopy without image intensification shall not be used.
(12) Hand-held Portable Dental X-ray Systems.
(a) X-ray equipment designed to be hand-held shall comply with Section R313-28-31, excluding Subsection R313-28-31(5), and with Section R313-28-80, excluding Subsections R313-28-80(7)(b) and R313-28-80(11)(b).
(b) Protective shielding of at least 0.5 millimeter lead equivalence shall be provided for the operator to protect the operator's torso, hands, face, and gonads from backscatter radiation. If the protective shielding is a backscatter shield attached to the x-ray unit, the shield shall be positioned as close to the patient as possible and the operator shall take care to remain in a protective position.
(c) Portable radiation machines designed to be hand-held are exempt from Subsection R313-28-35(7). The portable radiation machines shall be held by the tube housing support or handle and shall be used in accordance with the manufacturer's operating procedures.
(d) In addition to the requirements of Subsection R313-28-350(1), each operator shall complete the training program supplied by the manufacturer prior to using the x-ray unit. Records of training shall be maintained on file for examination by an authorized representative of the Director.