Utah Administrative Code (Current through November 1, 2019) |
R313. Environmental Quality, Waste Management and Radiation Control, Radiation |
R313-28. Use of X-Rays in the Healing Arts |
R313-28-57. Radiographic Systems Other Than Fluoroscopic, or Dental Intraoral Systems -- mA/mAs Linearity
-
The following requirements apply when the equipment is operated on a power supply as specified by the manufacturer for fixed x-ray tube potentials within the range of 40 percent to 100 percent of the maximum rated potentials.
(1) Equipment having independent selection of x-ray tube current, mA. Where the tube current is continuous, the average ratios of exposure to the indicated milliampere-seconds product, C/kg/mAs or mR/mAs, obtained at two consecutive tube current settings or at two settings differing by no more than a factor of two shall not differ by more than 0.10 times their sum.
(2) Equipment having a combined x-ray tube current-exposure time product, mAs, selector, but not a separate tube current, mA, selector. Where the tube current is continuous, the average ratios of exposure to the indicated milliampere-seconds product, C/kg/mAs or mR/mAs, obtained at two consecutive milliampere-seconds settings or at two settings differing by no more than a factor of two shall not differ by more than 0.10 times their sum.