R313. Environmental Quality, Waste Management and Radiation Control, Radiation |
![]() |
R313-28. Use of X-Rays in the Healing Arts |
R313-28-56. Radiographic Systems Other Than Fluoroscopic, or Dental Intraoral Systems -- Accuracy
Latest version.
-
Deviation of measured technique factors from indicated values of kVp and exposure time shall not exceed the limits specified for that system by its manufacturer. In the absence of manufacturer's specifications, the deviation shall not exceed ten percent of the indicated value for kVp and ten percent of the indicated value for times greater than 50 milliseconds.