R313-28-450. Minimum Design Requirements for an X-ray Machine Operator's Booth - New Installations Only
-
(1) Space requirements:
(a) The operator shall be allotted not less than 0.70 square meter (7.5 square feet) of unobstructed floor space in the booth.
(b) The minimum space as indicated above may be geometric configurations with no dimension of less than 0.61 meters (two feet).
(c) The space shall be allotted excluding encumbrances by the console, for example, overhang or cables, or other similar encroachments.
(d) The booth shall be located or constructed to ensure that unattenuated primary beam scatter originating on the examination table or at the wall mounted image receptor will not reach the operator's position in the booth.
(2) Structural Requirements.
(a) The booth walls shall be permanently fixed barriers of at least 2.13 meters (seven feet) high.
(b) When a door or movable panel is used as an integral part of the booth shielding, it must have a permissive device which will prevent an exposure when the door or panel is not closed.
(c) Shielding shall be provided to meet the requirements of R313-15.
(3) X-Ray Exposure Control Placement: The x-ray exposure control for the system shall be fixed within the booth and:
(a) shall be at least one meter (40 inches) from points subject to primary beam scatter, leakage or primary beam radiation; and
(b) shall allow the operator to use the majority of the available viewing windows.
(4) Viewing system requirements:
(a) When the viewing system is a window:
(i) the viewing window shall have a visible area of at least 0.09 square meters (one square foot);
(ii) regardless of size or shape, at least 0.09 square meters (one square foot) of the window area must be centered no less than 0.6 meters (two feet) from the open edge of the booth and no less than 1.5 meters (five feet) from the floor; and
(iii) the window shall have at least the same lead equivalence of that required in the booth's wall in which it is mounted.
(b) When the viewing system is by mirrors, the mirrors shall be so located as to accomplish the general requirements of R313-28-450(4)(a).
(c) When the viewing system is by electronic means:
(i) the camera shall be so located as to accomplish the general requirements of R313-28-450(4)(a); and
(ii) there shall be an alternate viewing system as a backup for the primary system.