R156-46a-502c. Calibration of Technical Instruments  


Latest version.
  • The requirement in Subsection 58-46a-303(3)(c) for calibration of all appropriate technical instruments used in practice is defined, clarified, and established as follows:

    (1) any audiometer used in the fitting of hearing aids shall be calibrated when necessary, but not less than annually;

    (2) the calibration shall include to ANSI standards calibration of frequency accuracy, acoustic output, attenuator linearity, and harmonic distortion; and

    (3) calibration shall be accomplished by the manufacturer, or a properly trained person, or an institution of higher learning equipped with proper instruments for calibration of an audiometer.