R313-35-120. X-Ray Systems Less than 1 MeV used for Non-Destructive Testing  


Latest version.
  • (1) Cabinet x-ray systems.

    Cabinet x-ray systems shall meet the requirements of R313-35-80.

    (2) Fixed Gauges.

    (a) Warning Devices. A light, which is clearly visible from all accessible areas around the x-ray system, shall indicate when the x-ray system is operating.

    (b) Personnel Monitoring. Notwithstanding R313-15-502(1)(a), individuals conducting x-ray system maintenance requiring the x-ray beam to be on shall be provided with and required to wear personnel monitoring devices.

    (3) Industrial and Other X-ray Systems.

    (a) Equipment.

    (i) The registrant shall perform visual and operability checks of indication lights and warning lights before use on each day the equipment is to be used to ensure that the equipment is in good working condition. If equipment problems are found, the equipment shall be removed from service until repaired.

    (ii) Inspection and routine maintenance of x-ray systems, interlocks, indication lights, exposure switches, and cables shall be made at intervals not to exceed six months or before the first use thereafter to ensure the proper functioning of components important to safety. If equipment problems are found, the equipment shall be removed from service until repaired.

    (iii) Records demonstrating compliance with R313-35-120(3)(a)(i) shall be made when problems with the equipment are found. These records shall be maintained for a period of three years.

    (iv) Records demonstrating compliance with R313-35-120(3)(a)(ii) shall be made. These records shall be maintained for a period of three years.

    (b) Controls. X-ray systems which produce a high radiation area shall be controlled to meet the requirements of R313-15-601.

    (c) Personnel Monitoring Requirements.

    (i) Registrants shall not permit individuals to conduct x-ray operations unless all of the following conditions are met.

    (A) Individuals shall wear a thermoluminescent dosimeter or film badge.

    (I) Each film badge or thermoluminescent dosimeter shall be assigned to and worn by only one individual.

    (II) Film badges shall be replaced at periods not to exceed one month and thermoluminescent dosimeters shall be replaced at periods not to exceed three months.

    (B) Individuals shall wear a direct reading dosimeter if conducting non-destructive testing at a temporary job site or in a room or building not meeting the requirements of R313-15-301.

    (I) Pocket dosimeters shall have a range from zero to two millisieverts (200 millirem) and must be recharged at the beginning of each shift.

    (II) Direct reading dosimeters shall be read and the exposures recorded at the beginning and end of each shift. Records shall be maintained for three years after the record is made.

    (III) Direct reading dosimeters shall be checked at intervals not to exceed 12 months for correct response to radiation and the results shall be recorded. Records shall be maintained for a period three years from the date the record is made. Acceptable dosimeters shall read within plus or minus 20 percent of the true radiation exposure.

    (IV) If an individual's ion-chamber pocket dosimeter is found to be off scale or if the individual's electronic personnel dosimeter reads greater than 2 millisieverts (200 millirems), and the possibility of radiation exposure cannot be ruled out as the cause, the individual's film badge or thermoluminescent dosimeter shall be sent for processing within 24 hours. In addition, the individual shall not resume work with sources of radiation until a determination of the individual's radiation exposure has been made.

    (d) Controls. In addition to the requirements of R313-15-601, barriers, temporary or otherwise, and pathways leading to high radiation areas shall be identified in accordance with R313-15-902.

    (e) Surveillance. During non-destructive testing applications conducted at a temporary job site or in a room or building not meeting the requirements of R313-15-301, the operator shall maintain continuous direct visual surveillance of the operation to protect against unauthorized entry into a high radiation area.